The CSS 225 is a simple holder for 225 mm long pH, ORP, conductivity, or dissolved oxygen sensors.
The CSS 225 static holder enables quick and easy installation of 225 mm long pH/ORP, conductivity, or dissolved oxygen sensors in process applications. The holder is designed for new applications and direct retrofit of existing sensors.
The CSS 225 static holder is part of a three-part system – sensor, holder, and cable. Users can remove M4 Knick sensors from the holder. The replacement cost of the system is less than those with integrated “potted “sensor and holder configurations. In addition, the holder has various threaded and flanged process fittings options. This helps to lower inventory costs.Mounting options: | ¾”, 1″, and 1½” MNPT Flange from 1″ ANSI |
Sensor length: | 225 mm |
Sensor connection: | Pg 13.5 |
Materials of construction: | 316L SS, PP, PVDF, other materials available upon request |
Operating pressure range: | 316L SS: 0…145 psi (0…10 bar) PP: 0 … 60 psi (0 … 4 bar), 14 … 50°F (-10°C …10°C) 60 psi (4.13 bar), 50 … 140°F (10 … 60°C) Ambient, 140 … 176°F (60 … 80°C) PVDF: 0 … 90 psi (0 … 6.21 bar), 14 … 32°F (-10 … 0°C) 90 psi (6.21 bar), 32 …194°F (0 … 90°C) Ambient , 194 …284°F (90 …140°C) |
Operating temperature range: | 316L SS: 14 … 284°F (-10 … 140°C) PP: 14 …176°F (-10 … 80°C) PVDF: 14 … 248°F (-10 … 120°C) |
Sealing materials: | EPDM, Viton™, Kalrez |
Protective cage: | Standard |
Insertion lengths: | 3.45″, 5.50″, 7.00″ Custom lengths on request |